{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65577",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide), Puerto Rico and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Czech Republic, Denmark (including Faroe Islands), Ecuador, Finland, France (including Martinique and Reunion Islands), Germany, Greece, Iceland, India, Ireland, Isreal, Italy, Luxemburg, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweded, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1677-2013",
      "product_description": "Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion.    The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or subcutaneously for post-operative pain management.",
      "product_quantity": "Code 2C1008KP: 56,400 units; Code 2C1009KP: 180,004 units; Code 2C1063KP: 154,937 units; Code 2C1087KP: 43,521 units; Code 2C1156KP: 7,908 units",
      "reason_for_recall": "There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage",
      "recall_initiation_date": "20090731",
      "center_classification_date": "20130710",
      "termination_date": "20160308",
      "report_date": "20130717",
      "code_info": "*** 1) Product Code: 2C1008KP; Lots: 08M036, 08N041, 09A030, 09B044, 09C053, 09D018, 09E019, 09F005;   *** 2) Product Code: 2C1009KP; Lots: 08M016, 08M035, 08M066, 09A010, 09A052, 09A065, 09A080, 09B038, 09B048, 09B060, 09B074, 09C03, 09C016, 09C044, 09C060, 09D011, 09D042, 09D049, 09D054, 09E021, 09E043, 09F002;   *** 3) Product Code: 2C1063KP; Lots: 08K078, 08M014, 08M060, 08N022, 08N053, 08N074, 09A032, 09A057, 09B006, 09B014, 09C007, 09C017, 09C054, 09C077, 09D015, 09D041, 09E012,09E018, 09E022, 09E044;   *** 4) Product Code: 2C1087KP; Lots: 08K074, 08M050, 08N051, 09A053, 09B063, 09C041, 09D044, 09E073;   *** 5) Product Code: 2C1156KP; Lots: 08M063, 09C042"
    }
  ]
}