  "results": [
    {
      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90681",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.",
      "recall_number": "Z-1675-2022",
      "product_description": "VITROS XT 3400 Chemistry System with Version 3.7.2 and below  Product Code: 6844458",
      "product_quantity": "506 units",
      "reason_for_recall": "Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.",
      "recall_initiation_date": "20220719",
      "center_classification_date": "20220902",
      "report_date": "20220914",
      "code_info": "UDI-DI: 10758750031986   Systems start with 3400XXXX where XXXX is unique for each analyzer"
    }
  ]
}