  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "800 MacArthur Blvd",
      "reason_for_recall": "Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor.  Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).",
      "address_2": "",
      "product_quantity": "4 units",
      "code_info": "Part No: 0998-00-0708-01  Software version 8.9",
      "center_classification_date": "20150526",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the state of TX.",
      "state": "NJ",
      "product_description": "Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm.  Part number 0998-00-0708-01.      The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "recall_number": "Z-1675-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "70959",
      "termination_date": "20170328",
      "more_code_info": "",
      "recall_initiation_date": "20150312",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}