  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "700 Orthopaedic Dr",
      "reason_for_recall": "The LCS COMPLETE RPS Knee System has higher rates of revision in Australia when the native patella is not resurfaced when compared to other unresurfaced PS implants.  If the LCS COMPLETE RPS Knee System is implanted and the native patella is not resurfaced, the patient may experience patello-femoral pain, potentially requiring a secondary procedure.",
      "address_2": "",
      "product_quantity": "28,732",
      "code_info": "All Lots.  Catalog numbers:  129416110  129416112  129416115  129416117  129416120  129416122  129416125  129416210  129416212  129416215  129416217  129416220  129416222  129416225  129416310  129416312  129416315  129416317  129416320  129416322  129416325  129416410  129416412  129416415  129416417  129416420  129416422  129416425  129416510  129416512  129416515  129416517  129416520  129416522  129416525  129416610  129416612  129416615  129416617  129416620  129416622  129416625  129416710  129416712  129416715  129416717  129416720  129416722  129416725  129417110  129417112  129417115  129417117  129417120  129417122  129417125  129417210  129417212  129417215  129417217  129417220  129417222  129417225  129417310  129417312  129417315  129417317  129417320  129417322  129417325  129417410  129417412  129417415  129417417  129417420  129417422  129417425  129417510  129417512  129417515  129417517  129417520  129417522  129417525  129417610  129417612  129417615  129417617  129417620  129417622  129417625  129417710  129417712  129417715  129417717  129417720  129417722  129417725  129495410  129495412  129495415  129495417  129495420  129495510  129495512  129495515  129495517  129495520  129495610  129495612  129495615  129495617  129495620",
      "center_classification_date": "20150526",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  CANADA, AUSTRALIA, AUSTRIA, BELGIUM, UK, FRANCE, GREECE, IRELAND, ISRAEL, ITALY, KOREA, LUXEMBOURG, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, SOUTH AFRICA, SWITZERLAND, THAILAND.",
      "state": "IN",
      "product_description": "LCS COMPLETE RPS inserts, various sizes. Knee prosthesis component for orthopedic surgery.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "recall_number": "Z-1673-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "71085",
      "termination_date": "20151030",
      "more_code_info": "",
      "recall_initiation_date": "20150304",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}