      "status": "Terminated",
      "city": "Warrendale",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68293",
      "recalling_firm": "Bayer Corp",
      "address_1": "100 Global View Dr",
      "address_2": "",
      "postal_code": "15086-7601",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the countries of  Australia, Brazil, Belgium, Canada,  Germany, Singapore, South Africa, Poland, Mexico and UAE.",
      "recall_number": "Z-1673-2014",
      "product_description": "Medrad Continuum MR Infusion system- non-wireless system    The MEDRAD Continuum MR Infusion System is designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR procedure functioning while either stationary or mobile. It is not intended to provide long-term patient care outside the MR environment. The system is to be used by trained medical staff, primarily critical care, emergency room and radiology nursing staff.",
      "product_quantity": "1347",
      "reason_for_recall": "There is a potential safety risk associated with potentially damaged locking pins within the bracket of non-wireless versions of Continuum systems shipped prior to March 2008.",
      "recall_initiation_date": "20140403",
      "center_classification_date": "20140530",
      "termination_date": "20150416",
      "report_date": "20140611",
      "code_info": "catalog number 3008895 : material numbers 3008236, 3009966, 3009917 with serial numbers: 1048  200619  200585  200656  300010  100066  14498  100087  200305  200663  200248  200606  300047  1040  100182  100187  200442  200037  200716  200111  300051  1041  100428  100128  200479  200586  200648  200009  300045  100107  100037  100213  200665  200034  200722  200134  300027  100272  100457  100215  200601  200109  200298  200135  100220  100273  100482  100141  200615  200121  200558  200593  300036  100401  11210  100488  200117  200313  200618  200465  300033  100409  10328  100129  200540  200515  200187  1045  300034  100404  15187  100083  200401  200119  20398  200045  300042  100299  10407  100006  200115  1068  200051  14857  300026  100300  100233  100188  200432  200267  1073  200557  300049  100399  100266  100226  200497  200689  200022  200322  30020  100400  14076  100271  200200  200771  200089  200346  300013  100184  100265  100069  200012  200075  200090  200655  300012  100185  1033  1047  200706  200076  200091  200509  300000  100405  100316  100098  200707  1047  200344  1006  200004  100173  100356  100051  200493  200031  200337  200175  300001  100174  10810  100148  200419  200512  200644  200103  300002  100259  100144  100297  200597  200430  200579  200436  300003  100283  100105  100449  100122  200348  200580  200104  300028  100375  14163  100452  200393  200276  200463  200105  300055  1000  11159  100454  200392  200205  200079  200204  300032  100517  100027  100458  200674  200469  200328  200443  300031  100518  100279  100072  200193  200268  200040  200444  300044  100521  100139  100084  200324  200000  200030  200225  300015  100212  100175  11132  200289  200627  200395  200421  1021  100284  14246  100304  200415  200583  200388  200396  300037  100036  100443  100257  200416  200087  200682  200047  300025  100545  100492  100447  200448  200136  200164  1020  300016  100433  100497  100529  200482  200137  200778  1012  300017  100434  100170  100487  200483  200270  200568  200124  300019  1019  100019  100537  200574  200581  200185  1008  300006  100296  14464  100515  200587  200641  200043  1064  300029  100135  10387  10213  200598  200213  200160  200049  300062  1038  14846  11179  200614  200186  200161  200450  300043  100354  100342  11310  200636  200157  200546  200113  300018  100408  100193  15065  200635  200233  200547  14877  300007  1031  100224  100210  200640  200700  200749  200237  300011  100199  100417  100243  200680  200514  14832  200142  300030  100473  100335  10509  200055  200352  14844  200252  14737  1027  100294  100336  200056  200318  200577  200671  300005  100438  100232  100484  200057  200221  200422  200397  300038  100439  100361  1030  200058  200149  200159  200709  300040  100033  100150  100161  200059  200211  1004  200710  300039  1002  100151  10354  200060  200414  200429  200166  300021  100197  100116  100359  200065  200708  200112  200165  300048  100198  100115  100372  200066  20020  200329  200084  300050  100260  100117  14709  200067  200020  200471  200220  300053  100510  100049  100057  200068  200254  200687  200378  300054  100511  100025  100054  200069  200223  200685  200561  300009  100512  100125  100501  100520  200150  1046  1016  300022  1011  100202  100496  200788  200226  1024  200464  100061  100203  100327  200351  200445  200338  200070  1017  100664  100029  200426  200418  200152  1023  100261  100665  100112  200477  200545  200138  200764  1001  100666  100113  200486  200144  200391  200294  100229  100667  11256  200011  200612  200454  200048  1037  100669  100114  200658  200478  200269  200455  1023  100468  11281  200326  200293  200386  200354  100096  200183  100411  200339  200460  200052  200170  100386  10320  100240  200340  200073  200244  200642  100134  100074  100008  200366  200074  200044  200053  100398  100132  100429  200694  200519  200470  200552  100060  100532  100093  200695  200209  200563  200697  100541  200498  10771  14906  200061  200467  200177  100253  100516  100362  14911  200370  200210  200350  100121  14378  100291  1032  200013  200622  1000  100159  100222  100396  1034  1031  200239  200250  100201  11015  100136  200502  1050  200434  200678  100045  100231  100485  200668  1049  200258  200253  100063  100250  100437  200439  200698  200259  200180  100277  100504  100474  200359  200088  200390  200171  100312  100506  100505  200092  200050  200217  200172  100313  100508  100467  200014  1007  200071  200173  100097  100419  100164  1027  200064  1061  200620  100255  100090  100481  200215  200628  200666  200623  100095  100357  100131  1070  200713  1062  200626  100326  100024  100479  200128  200334  200369  200630  100245  100491  100118  200361  200603  200371  200631  100341  100385  100058  200162  200300  200645  200632  100436  100004  14406  200231  200292  200127  200310  200495  100308  100502  200657  200555  200280  14644  100526  100166  100330  200141  1058  200041  200730  100332  100194  100337  200461  1060  200277  200468  100435  100079  100472  200659  200423  200368  200590  100427  200076-1  100378  1053  200301  200299  200646  100285  200076-2  100147  200094  200531  200664  200485  100047  15038  100048  1071  200224  200654  1011  100155  100012  11343  200408  200428  200452  200651  100163  100013  10282  20491  200380  1018  200721  1003  100137  100169  200473  200551  1036  200325  100207  100230  200156A  200661  200650  1037  200494  100195  100192  14614  200169  200385  200218  200007  100410  100130  100178  200570  200510  200394  200677  100387  200036  100413  200571  200576  200550  200528  100274  100176  15290  20598  200624  200275  200639  100244  100298  200699  300021  1054  200616  200582  100494  100464  100475  200353  200096  200146  200732  100440  100542  100206  200046  200097  200621  200145  10089  100016  100011  200120  10783  200367  200440  10091  100456  100317  200130  200108  200516  200319  100100  100041  100320  200357  200625  200327  200320  100156  100204  100522  200129  200564  200676  200181  100442  100258  100000  200679  10471  200525  200182  100026  11382  100018  200629  200530  200198  200693  100225  10660  100149  200072  200098  200675  200667  100275  14417  10997  200122  200435  200099  200126  100228  100290  14255  200522  200114  200717  200500  100524  100208  100160  200191  200333  200032  200365  100310  14385  1046  200376  200711  200273  200021  100426  100189  100288  200511  200712  200274  200360  100252  100480  100209  200513  200214  100293  1059  100420  100177  200063  200517  200228  200282  200362  100421  100104  100133  200518  200229  200039  200670  100328  100251  11170  200521  10328  200078  200246  100247  100146  11254  200131  200132  200194  200358  100038  1034  14117  200720  200316  200083  200317  100234  10371  14588  200377  200085  200672  10380  1010  100350  15386  200381  200431  1022  1074  100246  100360  15408  200456  1063  200207  200302  100425  1008  100364  200774  200110  1051  200303  100062  100183  100368  200775  200638  200331  200433  100196  100500  100369  200315  200262  200295  200255  100323  200336  100370  200492  200405  14935  200256  100489  100190  100366  200347  200596  200529  200155  100423  100307  100371  200278  200260  200265  100034  100392  100461  200199  200240  200264  1035  14072  14725  1026  200116  200026  100333  100168  100289  200195  200219  200481  1021  100483  100295  200438  200232  200772  1004  100186  200019  200541  200321  200643  1015  100270  100531  200633  200447  1076  100324  100126  10579  1041  200143  200496  100158  10342  100162  1075  200176  200372  100256  10317  100355  200556  200024  20016  100379  100138  14602  200537  20024  200345  100377  100140  14814  200535  200266  200683  100381  100772  100101  200148  200480  1077  100382  100453  100180  200503  1005  200565  100383  100448  100050  200609  1052  200008  100445  100450  100219  200272  200332  200524  100446  100451  100305  200523  1069  200139  100014  14903  1044  200167  200382  200409  100099  100380  11356  200188  200140  200578  100309  100067  100249  200189  200201  200662  100348  100441  100120  200190  200605  200399  100236  100040  100241  1033  200562  200222  100493  10674  100068  200095  200567  200508  100507  100363  100075  200548  200569  200086  1024  100569  100519  200575  200271  200446  100030  1057  100301  200673  200854  200501  100039  100227  100119  200686  200704  200458  100109  100314  100302  200227  200466  200459  1018  100315  100002  1014  200001  1067  1022  100172  100076  200077  200560  1055  100403  100465  100077  200407  200652  200003  100418  100466  100078  200035  1042  200684  100455  100469  100339  200404  200123  200599  15159  100470  100340  200283  200081  200602  100103  100153  100123  200284  200341  200208  100052  100154  200765  200311  200726  200400  100053  200297  100181  200312  200791  200153  100055  100431  10011  200798  200411  200238  100056  100432  100216  200799  200080  200705  100238  100143  200174  200290  200406  200323  14897  100070  100001  200592  200309  200715  100478  100459  100005  200286  200474  200364  100536  100223  15194  200287  200554  200133  100064  100393  100444  200192  1017  200288  1012  100278  100476  200308  1001  200356  1026  1045  100142  200330  200507  200462  1030-1  100017  100009  200118  200196  200538  100322  100422  100544  200499  200457  200082  100390  100292  14414  200504  200242  200343  100394  100102  100373  200505  1019  200532  100085  100022  1056  200506  200151  200389  100407  100023  14471  1065  200660  1072  100059  100015  1016  200617  200637  200147  100157  100071  100073  200106  200534  1039  100082  100530  200156  200203  200236  200197  100094  100044  100248  1029  200029  200038  100262  10594  15097  200335  200544  200375  100263  14650  100127  200033  200216  200158  100264  15236  14380  200093  200263  200005  100088  200042  100534  200604  200027  200427  100334  1028  1066  200476  200475  200355  1007  1009  100106  200212  200387  1015  15739  10567  100028  200154  200004  200125  100179  100220  100167  200600  200412  200572  100242  14381  200669  200417  200549  100269  100003  200594  200235  200028  100430  100607  200015  200243  200062  100525  100191  200384  200010  1038  200425  100325  200403  200608  200363  100329  100124  200202  200584  200653  100145  100217",
      "more_code_info": ""
    }
  ]
}