  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "700 Orthopaedic Dr",
      "reason_for_recall": "The LCS COMPLETE RPS Knee System has higher rates of revision in Australia when the native patella is not resurfaced when compared to other unresurfaced PS implants.  If the LCS COMPLETE RPS Knee System is implanted and the native patella is not resurfaced, the patient may experience patello-femoral pain, potentially requiring a secondary procedure.",
      "address_2": "",
      "product_quantity": "19,059",
      "code_info": "All Lots.  Catalog numbers:  129411010  129411020  129411030  129411040  129411050  129411060  129411070  129412010  129412020  129412030  129412040  129412050  129412060  129412070  129495040  129495050  129495060  129496040  129496050  129496060",
      "center_classification_date": "20150526",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  CANADA, AUSTRALIA, AUSTRIA, BELGIUM, UK, FRANCE, GREECE, IRELAND, ISRAEL, ITALY, KOREA, LUXEMBOURG, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, SOUTH AFRICA, SWITZERLAND, THAILAND.",
      "state": "IN",
      "product_description": "LCS COMPLETE RPS Femoral Implants, various sizes and orientation.  Knee prosthesis component for orthopedic surgery.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "recall_number": "Z-1672-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "71085",
      "termination_date": "20151030",
      "more_code_info": "",
      "recall_initiation_date": "20150304",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}