  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87594",
      "recalling_firm": "Bard Access Systems Inc.",
      "address_1": "605 N 5600 W",
      "address_2": "N/A",
      "postal_code": "84116-3738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN,  TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.",
      "recall_number": "Z-1668-2021",
      "product_description": "REF 7617405, Groshong NXT ClearVue Catheter Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741086998 - Product Usage: is designed for use when central venous catheterization is prescribed.",
      "product_quantity": "N/A",
      "reason_for_recall": "Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector  and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock",
      "recall_initiation_date": "20210407",
      "center_classification_date": "20210520",
      "termination_date": "20230601",
      "report_date": "20210526",
      "code_info": "Lot No. reds4650  redt0780  redt1528  redt4250  redu0070  redu0706  redu1526  REDU2456  redv1194  REDV1893  redv1894  redv2160  REDV2537  redv2538  redv2539  redv2924  redv4016  REDW0575  redw0837  redw4033  redw4034  redw4035  redw4036  redw4037  redw4038  REDW4039  redw4040  redw4042  redw4164  redw4170  redx0604  redx0743  redx0972  redx0973  redx0974  redx0975  redx0976  redx0977  redx0978  redx0979  redx0980  redx0991  redx0992  redx0993  redx0994  redx0995  redx0996  redx1339  redx1340  redx1504  redx2371  redx2483  redx4877  redy0094  redy0106  redy0107  redy0662  redy0713  redy1680  REDY2962  redz0056  redz0185  redz0664  redz0697  redz1649  redz1710  REEN1131  reen1240  reen1911  reen1982"
    }
  ]
}