      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79505",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Canada, Mexico, and Thailand",
      "recall_number": "Z-1668-2018",
      "product_description": "Dimension¿ Thyroid Stimulating Hormone (TSHL) Flex¿ reagent cartridge",
      "product_quantity": "N/A",
      "reason_for_recall": "The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.",
      "recall_initiation_date": "20180102",
      "center_classification_date": "20180504",
      "termination_date": "20230210",
      "report_date": "20180516",
      "code_info": "TSHL, radioimmunoassay, thyroid-stimulating hormone, RF612, SMN# 10464524    Lot codes:  EB7276  EC7298  EA7318  EB7346  EC7346  EA8012  EA8033  EA8051  EC8080  EB8107  EB8121  EA8143  EB8143  EA8177  EB8177  EB8199  EA8223  EA8241  EA8261"
    }
  ]
}