      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "Study and all acquired images deleted when using systems with software version VE10E.  Acquisitions have to be repeated.  Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery.  Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.",
      "address_2": "",
      "product_quantity": "133 total",
      "code_info": "Model Number of device   10762471with serial numbers: 5043  5085  5069  5086  5037  5072  5049  5039  5120",
      "center_classification_date": "20150523",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "PA",
      "product_description": "SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1668-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71115",
      "termination_date": "20170320",
      "more_code_info": "",
      "recall_initiation_date": "20150413",
      "postal_code": "19355-1406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}