  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "Study and all acquired images deleted when using systems with software version VE10E.  Acquisitions have to be repeated.  Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery.  Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.",
      "address_2": "",
      "product_quantity": "133 total",
      "code_info": "Model Number of device   10762470, with serial numbers: 24077  24051  24038  24116  24082  24033  24136  24155  24159  24066  24106  24052  24128  24059  24048  24099  24100  24028  24046  24063  24083  24135  24131  24091  24024  24129  24103  24148  24142  24014  24015  24026  24109  24034  24108  24049  24058  24113  24154  24156  24039  24081  24114  24090  24105  24075  24016  24025  24042  24037  24053  24121  24087",
      "center_classification_date": "20150523",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "PA",
      "product_description": "SIEMENS Ysio Max;  a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting    x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1667-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71115",
      "termination_date": "20170320",
      "more_code_info": "",
      "recall_initiation_date": "20150413",
      "postal_code": "19355-1406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}