  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68281",
      "recalling_firm": "Beaver-Visitec International Inc.",
      "address_1": "411 Waverley Oaks Rd Ste 229",
      "address_2": "",
      "postal_code": "02452-8422",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AL, CT, IN, NC, NY, TX, and VA.",
      "recall_number": "Z-1664-2014",
      "product_description": "Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in)  REF 585006    Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate. These products are used in all sections of the eye. This cannula is a general use cannula mainly designed for the precise maintenance of optimum anterior chamber depth",
      "product_quantity": "9970 units",
      "reason_for_recall": "Contains a sharp tip needle when the product should contain a blunt tip cannula",
      "recall_initiation_date": "20140506",
      "center_classification_date": "20140527",
      "termination_date": "20140905",
      "report_date": "20140604",
      "code_info": "Lot number: 3079324",
      "more_code_info": ""
    }
  ]
}