  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87594",
      "recalling_firm": "Bard Access Systems Inc.",
      "address_1": "605 N 5600 W",
      "address_2": "N/A",
      "postal_code": "84116-3738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN,  TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.",
      "recall_number": "Z-1661-2021",
      "product_description": "REF 9617408, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036009 - Product Usage: is designed for use when central venous catheterization is prescribed.",
      "product_quantity": "N/A",
      "reason_for_recall": "Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector  and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock",
      "recall_initiation_date": "20210407",
      "center_classification_date": "20210520",
      "termination_date": "20230601",
      "report_date": "20210526",
      "code_info": "Lot No. reep1720  redy0653  REEN1127  REEN1905  reeq0634  rees0249  reet2073  reev1213  reew4243"
    }
  ]
}