  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92006",
      "recalling_firm": "Maquet Medical Systems USA",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution",
      "recall_number": "Z-1659-2023",
      "product_description": "QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass  Model                                       Item No. HMO 70000-USA                    701067820 HMO 71000-USA                    701067823 BEQ-HMO 71000-USA           701067829",
      "product_quantity": "0 (US)",
      "reason_for_recall": "Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.",
      "recall_initiation_date": "20230518",
      "center_classification_date": "20230530",
      "report_date": "20230607",
      "code_info": "Model                                       Item No.            UDI HMO 70000-USA                    701067820         4058863019147 HMO 71000-USA                    701067823         4058863017341 BEQ-HMO 71000-USA           701067829         4058863017372  All lots"
    }
  ]
}