{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Drachten",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98473",
      "recalling_firm": "BD KIESTRA LAB AUTOMATION",
      "address_1": "Marconilaan 6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: KS, MD, MN, NJ, TX, VA;  International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;",
      "recall_number": "Z-1658-2026",
      "product_description": "BD Kiestra\" ReadA;  Catalog No.: 446948.",
      "product_quantity": "82 units (7 US, 65 OUS)",
      "reason_for_recall": "In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.",
      "recall_initiation_date": "20260206",
      "center_classification_date": "20260327",
      "report_date": "20260408",
      "code_info": "Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;",
      "more_code_info": ""
    }
  ]
}