{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64973",
      "recalling_firm": "The Anspach Effort, Inc.",
      "address_1": "4500 Riverside Drive",
      "address_2": "N/A",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide including the states of:  AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of:  Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;",
      "recall_number": "Z-1657-2013",
      "product_description": "***REF SC2100***eMax 2 Plus Console***SN G 10307653002***    Product Usage:  The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the \u001cFree running\u001d state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.",
      "product_quantity": "61 units",
      "reason_for_recall": "The RPM of the Motor Handpiece may be less than 80,000 RPM.",
      "recall_initiation_date": "20110624",
      "center_classification_date": "20130702",
      "termination_date": "20131224",
      "report_date": "20130710",
      "code_info": "SC2100 System Console Emax 2 Plus;  SC2101 System Console, Single Port with irrigation;  SC2102 System Console, Single Port, without Irrigation."
    }
  ]
}