  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87594",
      "recalling_firm": "Bard Access Systems Inc.",
      "address_1": "605 N 5600 W",
      "address_2": "N/A",
      "postal_code": "84116-3738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN,  TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.",
      "recall_number": "Z-1656-2021",
      "product_description": "REF 7655405, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035279 - Product Usage: is designed for use when central venous catheterization is prescribed.",
      "product_quantity": "N/A",
      "reason_for_recall": "Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector  and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock",
      "recall_initiation_date": "20210407",
      "center_classification_date": "20210520",
      "termination_date": "20230601",
      "report_date": "20210526",
      "code_info": "Lot No. reds2369  reds3369  reds4464  reds4588  redt0655  redt0787  redt1710  redt1760  redt2668  redt2909  redt4286  redt4309  redu0080  redu0745  redu0826  redu0827  redu1375  redu1533  redu2186  redu2403  redu2995  redu3073  redu3829  redu3920  redv1130  redv1285  REDV2022  redv2095  redv2812  redv2886  redv3944  REDV4092  REDW0350  REDW0460  redw1019  redw1151  redw1668  redw1669  redw1695  redw2471  redw2592  redw3754  redx0981  redx0982  redx0984  redx0985  redx0986  redx0987  redx1000  redx1001  REDX1002  redx1003  redx4676  redy0095  redy0096  redy0108  redy0109  redy0644  redy0703  REDZ0125  redz0433  redz1593  redz2836  redz3691  redz3692  redz3694  reen1167  reen1970  reen1971  reen3067  reen3286  reen4660  reen4733  reen5319  reep0814  reep1710  reeq2360  reeq2384  reeq2507  reeq2508  REDX3652  redy3370  REEQ2519"
    }
  ]
}