  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Clarksville",
      "address_1": "1865 Ashland City Rd Ste H",
      "reason_for_recall": "Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts.  Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.",
      "address_2": "",
      "product_quantity": "22 units",
      "code_info": "Serial numbers of serviced devices, serviced 09/2020-02-2021:  12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,",
      "center_classification_date": "20210601",
      "distribution_pattern": "Us distribution to the states of IN and TX.",
      "state": "TN",
      "product_description": "Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.",
      "report_date": "20210609",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "The Biomed Guys",
      "recall_number": "Z-1651-2021",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "87768",
      "more_code_info": "",
      "recall_initiation_date": "20210325",
      "postal_code": "37043-6455",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}