  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65207",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide including the states of  AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of:  Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain",
      "recall_number": "Z-1651-2013",
      "product_description": "ORTHO¿ Assay Software (OAS) Server, Catalog # 6842840    Usage:  The ORTHO¿ Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing.",
      "product_quantity": "31 units",
      "reason_for_recall": "There is evidence that an electrical short within the systems connector has caused an accumulation of  heat which has radiated back through the wiring harness causing the wires and connectors to melt.",
      "recall_initiation_date": "20130506",
      "center_classification_date": "20130701",
      "termination_date": "20180516",
      "report_date": "20130710",
      "code_info": "Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331"
    }
  ]
}