  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82703",
      "recalling_firm": "Maquet Cardiovascular, LLC",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1645-2019",
      "product_description": "Maquet Getinge-BEQ-TOP 250011/4 1/4 Less than 20 kg up  \t\t  Material: 701053906R01",
      "product_quantity": "N/A",
      "reason_for_recall": "Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured",
      "recall_initiation_date": "20190418",
      "center_classification_date": "20190530",
      "termination_date": "20250728",
      "report_date": "20190605",
      "code_info": "Batch Numbers:   3000068802  3000071398  3000074190  3000076543"
    }
  ]
}