  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients.  Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard.  Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application.  DB may lock under heavy and multitasks.",
      "address_2": "",
      "product_quantity": "75",
      "code_info": "SOMATOM Emotion 16 (2007)  10165977   with serial numbers:   86514  32490  80553  32552  32530  86406  86558  80542  80461  86405  32512  80441  86431  80467  86425  80442  86429  86557  80434  32581  32637  32597  80448  80431  32624  32469  80446  32578  80430",
      "center_classification_date": "20150522",
      "distribution_pattern": "Nationwide Distribution including Puerto Rico.",
      "state": "PA",
      "product_description": "SOMATOM Emotion 16, the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1645-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71101",
      "termination_date": "20160901",
      "more_code_info": "",
      "recall_initiation_date": "20150420",
      "postal_code": "19355-1406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}