  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67957",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA ( nationwide)  Distribution.",
      "recall_number": "Z-1645-2014",
      "product_description": "550 TxT Treatment Table of the Digital Linear Accelerator (LINAC)    To deliver X-ray radiation for therapeutic treatment of cancer.",
      "product_quantity": "120",
      "reason_for_recall": "Table may lose calibration during patient treatment, which may cause the possibility of mistreatment.",
      "recall_initiation_date": "20140401",
      "center_classification_date": "20140521",
      "termination_date": "20141211",
      "report_date": "20140528",
      "code_info": "material #8139789 , 5857912,  7360717, 1940035, 8139789,  4504200,  5863472, serial # 5377, 5087, 70-4262, 5170, 5282   5506   3287   5527   5517   70-4378   3695   5364   5060   5205   5688   5743   70-4283   5190   5157   5343   5844   5148   5391   5300   70-4296   5474   70-4363   70-4368   3793   3817   5833   5899   5207   5355   70-4159   3524   5281   5894   5374   5558   3631   3835   5784   2765   5367   70-4354   70-4356   5540   5488   5590   5657   5350   5821   3413   5222   5843   5179   70-4129   5851   5098   5153   5700   5410   3825   5865   5118   5548   3769   5861   70-4188   5748   70-4317   5737   5830   5385   5656   3778   5088   5665   5823   5353   5630   5480   5096   5572   3873   5417   4059   5500   5388   70-4379   5226   5601   5640   5033   5177   5707   5398   4091   5532   70-4306   70-4161   5424   5755   5326   5145   5699   4079   4082   70-4077   70-4113   5525   5419   3975   5093   5481   5396   3089   5095   5154"
    }
  ]
}