  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hazelwood",
      "state": "MO",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73413",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "address_2": "N/A",
      "postal_code": "63042-2320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Foreign distribution in the countries of Czech Republic, Hong Kong (China), Hungary, Japan, Poland, South Africa, Switzerland, and Turkey.",
      "recall_number": "Z-1643-2016",
      "product_description": "Vitek 2 Gram Positive Susceptibility Test Cards (AST-P642 REF 418604), 20 cards/kit.     Product Usage:  The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.",
      "product_quantity": "15,006 kits",
      "reason_for_recall": "Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.",
      "recall_initiation_date": "20160224",
      "center_classification_date": "20160511",
      "termination_date": "20180308",
      "report_date": "20160518",
      "code_info": "Lots 742343610 exp. 5/11/16, 742347220 exp. 6/16/16, 742350010 exp. 7/14/16, 742352820 exp. 8/11/16, 742355710 exp. 9/9/16, 742362540 exp. 11/16/16, 742364020 exp. 12/1/16, 742368210 exp. 1/12/16, 742376540 exp. 4/5/17, 742383610 exp. 6/15/17, and 742389220 exp. 8/10/17."
    }
  ]
}