{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rastatt",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79811",
      "recalling_firm": "Maquet Cardiopulmonary Ag",
      "address_1": "Kehler Strasse 31",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The product was distributed to the following US state: MO.  The product was distributed to the following foreign countries: France, Germany, Switzerland, and Thailand.",
      "recall_number": "Z-1642-2018",
      "product_description": "CARDIOHELP Gas Bottle Holder HKH 8880, Product Code/Part Number 70105.0901    Product Usage:  The Cardiohelp Gas Bottle (or Cylinder) Holder HKH 8880 is an accessory for the CARDIOHELP System and required for inter-hospital patient transport in aircraft and road vehicles (mobile intensive care units and ambulances). Inter-hospital patient transport is transport of the patient between clinics, where the clinical environment is left (e.g., transport in vehicles such as ambulances or aircraft). The gas cylinder holder serves to hold an oxygen cylinder in aircraft.",
      "product_quantity": "14 units (1 US, 13 International)",
      "reason_for_recall": "There is a possibility for the clamps of the gas bottle holder to crack or break when exposed to strong vibrations.",
      "recall_initiation_date": "20180404",
      "center_classification_date": "20180504",
      "termination_date": "20190329",
      "report_date": "20180516",
      "code_info": "US: 98880-020    Foreign: 98880 - 002, 98880 - 004, 98880 - 005, 98880 -  006, 98880-009, 98880-010, 98880-011, 98880-012, 98880-  013, 98880-014, 98880-016, 98880-017, 98880-018"
    }
  ]
}