  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98454",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.",
      "recall_number": "Z-1639-2026",
      "product_description": "Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm  Product Name: Everest Bipolar Cutting Forceps  Model/Catalog Number: 3006  Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.",
      "product_quantity": "164 units",
      "reason_for_recall": "Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.",
      "recall_initiation_date": "20260225",
      "center_classification_date": "20260325",
      "report_date": "20260401",
      "code_info": "Model:  3006; UDI:  00821925035898; Lot Numbers: All unexpired;",
      "more_code_info": ""
    }
  ]
}