{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Murrysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87753",
      "recalling_firm": "Philips Respironics, Inc.",
      "address_1": "1001 Murry Ridge Ln",
      "address_2": "N/A",
      "postal_code": "15668-8517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US nationwide to states of:  LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN.  Brazil, China, and Japan",
      "recall_number": "Z-1630-2021",
      "product_description": "BiPAP A40, Brazil: Continuous ventilator, non-life-supporting.  Material Number: BR1111169",
      "product_quantity": "4 units",
      "reason_for_recall": "Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.",
      "recall_initiation_date": "20210406",
      "center_classification_date": "20210521",
      "termination_date": "20240117",
      "report_date": "20210602",
      "code_info": "Serial Number: V306120599B37, V3061207124CF, V30612105753F, V306121267414"
    }
  ]
}