  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Englewood",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70946",
      "recalling_firm": "Baxter Corporation Englewood",
      "address_1": "014445 Grasslands Dr",
      "address_2": "N/A",
      "postal_code": "80112-7062",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-1628-2015",
      "product_description": "ABACUS SE (Single-Workstation Edition) and ABACUS ME (Multi-Workstation Edition)    Product Usage:  The ABACUS Software is a Windows-based order entry software application for comprehensive total parenteral nutrition (TPN) calculations and label printing.",
      "product_quantity": "882",
      "reason_for_recall": "Baxter Corporation is conducting a field action for the ABACUS SE and ME models due to the possibility that the compounder will load the incorrect formula upon scanning the bag label when two or more different formula files have been created with the same order number.",
      "recall_initiation_date": "20150408",
      "center_classification_date": "20150518",
      "termination_date": "20161201",
      "report_date": "20150527",
      "code_info": "Product Codes: 8300-0168 ,8300-0169 ,8300-0157 ,8300-0158 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047 ,8300-0046 ,8300-0047,"
    }
  ]
}