  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67923",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slovakia, Sweden, and Turkey.",
      "recall_number": "Z-1628-2014",
      "product_description": "Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors.   The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.",
      "product_quantity": "188,520 devices",
      "reason_for_recall": "The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.",
      "recall_initiation_date": "20140404",
      "center_classification_date": "20140516",
      "termination_date": "20171107",
      "report_date": "20140528",
      "code_info": "Patient Monitor Product numbers M3002A, M8102A, M8105A, M8001A, M8002A, M8003A, M8004A, M8027A, M8001AU, M8002AU, M8003AU, and M8004AU with software releases up to and including G.0."
    }
  ]
}