  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71167",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to the states of MN, FL & TX.",
      "recall_number": "Z-1623-2015",
      "product_description": "CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985.    Designed to be a head and whole body CT scanner utilizing a new solid state detector.",
      "product_quantity": "4 USA",
      "reason_for_recall": "GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res",
      "recall_initiation_date": "20150422",
      "center_classification_date": "20150515",
      "termination_date": "20150715",
      "report_date": "20150527",
      "code_info": "Mfg Lot or Serial # System ID  00000331100CN8 904223CT  00000345975CN7 409838LS16  00000337343CN8 512OCCT    00000344962CN6 INSIGHTPCT1356"
    }
  ]
}