  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85113",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "N/A",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, VI, RI, PR.    O.U.S.: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Gibraltar, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jersey, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Malta, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam",
      "recall_number": "Z-1622-2020",
      "product_description": "Philips V60 Ventilators with Power Management PCBA part number 1055906",
      "product_quantity": "23581",
      "reason_for_recall": "Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia",
      "recall_initiation_date": "20200305",
      "center_classification_date": "20200331",
      "report_date": "20200408",
      "code_info": "Serial Numbers:  100002908 to 100027093;  100027096 to 100085389;  100085399 to 100109746;  100109868 to 201003390;  201003392 to 201010952"
    }
  ]
}