      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65301",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.",
      "recall_number": "Z-1620-2013",
      "product_description": "Bigliani/Flatow¿ The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40.     Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.",
      "product_quantity": "722 units (includes 2010 market withdrawal)",
      "reason_for_recall": "This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included.  Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.",
      "recall_initiation_date": "20130523",
      "center_classification_date": "20130701",
      "termination_date": "20141218",
      "report_date": "20130710",
      "code_info": "Item Number: 00-4305-019-40 with Lot Numbers 61588143 and 61612276.  2010 Market Withdrawal: Item #:430501940  Lot #'s: 60768814, 60806551, 60893241, 60946820, 60999665, 61031668, 61110808, 61170678, 61205673, 61244194, 61392456, 61430127"
    }
  ]
}