{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "El Segundo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85089",
      "recalling_firm": "Karl Storz Endoscopy",
      "address_1": "2151 E Grand Ave",
      "address_2": "N/A",
      "postal_code": "90245-5017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and country of Canada.",
      "recall_number": "Z-1618-2020",
      "product_description": "Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers:    1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC      2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC.    Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.",
      "product_quantity": "27,165 scopes",
      "reason_for_recall": "The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.",
      "recall_initiation_date": "20191206",
      "center_classification_date": "20200331",
      "termination_date": "20220701",
      "report_date": "20200408",
      "code_info": "All Lots"
    }
  ]
}