  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Covington",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73827",
      "recalling_firm": "C.R. Bard, Inc.",
      "address_1": "8195 Industrial Blvd NE",
      "address_2": "N/A",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution to AL, AK, AZ, AR, CA, CO, CT, DEL, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and Internationally  to Argentina, Australia, Austria,  Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Finland, France, Germany, Gibraltar, Greece, Guam, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait,  Luxembourg, Malaysia, Mexico, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia,  Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab  Emirates, United Kingdom",
      "recall_number": "Z-1617-2016",
      "product_description": "Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V  The product is a thermal regulating system indicated for monitoring and controlling patient temperature",
      "product_quantity": "2613",
      "reason_for_recall": "The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM).  This premature coin cell battery drainage could render the device unresponsive upon System startup.",
      "recall_initiation_date": "20151109",
      "center_classification_date": "20160506",
      "termination_date": "20170606",
      "report_date": "20160518",
      "code_info": "Product Codes - 5000-00-00, 5000-01-01, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05"
    }
  ]
}