  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71045",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and Tobago; United States, including Puerto Rico; and Uruguay.",
      "recall_number": "Z-1616-2015",
      "product_description": "COULTER DxH Diluent, Catalog No. 628017, Part No. A59956.    For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.",
      "product_quantity": "829,888 units total (659,584 units in US)",
      "reason_for_recall": "Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.  The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.",
      "recall_initiation_date": "20150413",
      "center_classification_date": "20150514",
      "termination_date": "20151001",
      "report_date": "20150520",
      "code_info": "Lot No. 3099980 - 3512190, expiration dates: 17 April 2015 - 26 August 2016.",
      "more_code_info": ""
    }
  ]
}