  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.",
      "address_2": "",
      "product_quantity": "590,136 boxes",
      "code_info": "5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001",
      "center_classification_date": "20140513",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "IN",
      "product_description": "ACCU-CHEK Compact Plus Test Strips  For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip  51 count retail strip; 102 count retail strip; 51 count MedBen strip  51 count Mail Order strip; Sample Kit 16 per case.    The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.",
      "report_date": "20140521",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-1616-2014",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "67961",
      "termination_date": "20140918",
      "more_code_info": "",
      "recall_initiation_date": "20140414",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}