  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.",
      "address_2": "PO BOX 6101",
      "product_quantity": "17801",
      "code_info": "Model numbers: D045912, D193440, D140066;   Material Numbers: 10472176, 10636928, 10465412, 10461038, 10444828, 10444829, 10461849, 10461043, 10444837, 10444838",
      "center_classification_date": "20150514",
      "distribution_pattern": "Distributed in the country of Japan.",
      "state": "DE",
      "product_description": "Dimension EXL with LM, Dimension EXL 200, Dimension RxL, Dimension RxL HM, Dimension RxL Max, Dimension RxL Max HM, Dimension Xpand, Dimension Xpand HM, Dimension Xpand Plus, Dimension Xpand Plus HM.         The Dimension EXL clinical chemistry system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.",
      "report_date": "20150520",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-1615-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70913",
      "termination_date": "20160817",
      "more_code_info": "",
      "recall_initiation_date": "20150402",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}