  "results": [
    {
      "status": "Ongoing",
      "city": "Northridge",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87476",
      "recalling_firm": "Medtronic Inc.",
      "address_1": "18000 Devonshire St",
      "address_2": "N/A",
      "postal_code": "91325-1219",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-1614-2021",
      "product_description": "Medtronic REF: MMT-7020D3 Guardian Sensor (3)  for the following packaging:     GS3 1 PK OUS3/ GTIN:\t00643169704121  GS3 1PK 9L OUS3/ GTIN: 00763000318895  Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.    *Product not distributed within US.",
      "product_quantity": "13,014 units",
      "reason_for_recall": "Use of the continuous glucose monitoring system while actively taking   hydroxyurea, an anti-neoplastic drug  could result in inaccurate elevated sensor glucose readings.",
      "recall_initiation_date": "20210226",
      "center_classification_date": "20210517",
      "report_date": "20210526",
      "code_info": "All Lot numbers within active shelf life"
    }
  ]
}