  "results": [
    {
      "status": "Ongoing",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92090",
      "recalling_firm": "TELEFLEX LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-1611-2023",
      "product_description": "Teleflex (manufacturer: Vascular SOLUTIONS) D-Stat Flowable Hemostat, REF 4000; for sealing residual oozing of tissue tracts of femoral access sites that have been previously closed by suture/collagen-based hemostatic devices.",
      "product_quantity": "1140 units",
      "reason_for_recall": "The product contains an expired diluent vial. If the impacted product is used, the sterility of the diluent component cannot be assured as it has passed beyond the studied shelf life of the product.",
      "recall_initiation_date": "20230412",
      "center_classification_date": "20230518",
      "report_date": "20230524",
      "code_info": "UDI/DI M20640000, Lot Numbers:  717535, exp. 04/27/2023; 717413, exp. 04/25/2023"
    }
  ]
}