{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Round Lake",
      "address_1": "Route 120 And Wilson Rd",
      "reason_for_recall": "Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.",
      "address_2": "",
      "product_quantity": "601,230 total",
      "code_info": "Product Codes: 2C1071KJP, 2C1073KJP, 2C1075KJP, 2C1080KJP, 2C1082KJP, 2C1955KJP, 2C1976KJ.    Expiration Dates: 06/30/13, 12/31/13, 06/30/14, 12/31/14",
      "center_classification_date": "20130627",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.",
      "state": "IL",
      "product_description": "Brand Name:  Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor:    Single day Infusor 2 ml/h,   Single Day INFUSOR 2 ml/h System,   Half Day INFUSOR SV 5 ml/h System,   Two Day INFUSOR 2 ml/h System,   Two day Infusor 2 ml/h,   Multiday Infusor 0.5 ml/h,   Multiday INFUSOR 0.5ml/h System,   Seven Day Infusor 0.5 ml/h,   Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor       Device Listing: D018045    The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Baxter Healthcare Corp.",
      "recall_number": "Z-1610-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65364",
      "termination_date": "20140604",
      "more_code_info": "",
      "recall_initiation_date": "20130607",
      "postal_code": "60073-9799",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}