  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Round Lake",
      "address_1": "Route 120 And Wilson Rd",
      "reason_for_recall": "Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.",
      "address_2": "",
      "product_quantity": "601,230 total",
      "code_info": "Product Code: 2C1073KJP    Lot Numbers: 08M002, 08N007, 08N029, 08N030, 08N034, 08N069, 08N070, 09A001",
      "center_classification_date": "20130627",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.",
      "state": "IL",
      "product_description": "Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device       Device Listing: D020871    Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate.",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Baxter Healthcare Corp.",
      "recall_number": "Z-1609-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65364",
      "termination_date": "20140604",
      "more_code_info": "",
      "recall_initiation_date": "20130607",
      "postal_code": "60073-9799",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}