      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79783",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1606-2018",
      "product_description": "Mutlidiagnost-Eleva with Flat Detector 708034    As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.",
      "product_quantity": "206",
      "reason_for_recall": "The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.",
      "recall_initiation_date": "20180326",
      "center_classification_date": "20180503",
      "termination_date": "20200612",
      "report_date": "20180509",
      "code_info": "System Serial #    201  137  223  228  219  60  272  35  111  85  121  38  120  185  168  32  30  100  235  34  148  149  67  269  191  118  170  241  210  255  8  273  33  240  131  99  193  36  138  96  222  278  257  246  23  147  162  220  47  179  221  126  195  46  144  254  189  75  56  217  83  5  167  6  154  155  156  177  183  157  245  116  62  207  81  113  134  103  115  55  200  180  175  78  93  24  106  110  123  266  3  URF.173  14  202  208  247  225  28  216  262  9  176  261  259  174  114  68  63  122  135  205  224  39  98  80  22  251  142  253  84  181  264  163  73  94  95  160  199  206  159  112  196  151  25  238  150  4  117  50  58  51  79  249  119  204  133  18  13  15  16  17  129  244  227  234  130  31  12  27  11  128  171  237  76  1  143  248  109  42  43  44  172  213  212  2  48  127  19  82  69  70  65  242  54  194  136  190  152  105  233  71  97  132  197  53  211  72  187  209  41  146  125  268  26  77  40"
    }
  ]
}