  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Powell",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85102",
      "recalling_firm": "DeRoyal Industries Inc",
      "address_1": "200 Debusk Ln",
      "address_2": "N/A",
      "postal_code": "37849-4703",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "VA, CA, FL, AZ",
      "recall_number": "Z-1604-2020",
      "product_description": "DeRoyal SECURITY CUFFS, Single Strap, QR Buckle Cuff and Bed, UNIVERSAL, REF M8138 - Product Usage: patient restraint.",
      "product_quantity": "94 pairs",
      "reason_for_recall": "One lot of product was inadvertently manufactured with a buckle which does not meet the mechanical hold force required for this device.",
      "recall_initiation_date": "20200227",
      "center_classification_date": "20200330",
      "termination_date": "20220510",
      "report_date": "20200408",
      "code_info": "GTIN: 0749756946865  Lot Code: 50192186"
    }
  ]
}