  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79783",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1603-2018",
      "product_description": "MultiDim Eleva 708032      As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.",
      "product_quantity": "184",
      "reason_for_recall": "The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.",
      "recall_initiation_date": "20180326",
      "center_classification_date": "20180503",
      "termination_date": "20200612",
      "report_date": "20180509",
      "code_info": "System Serial #    32  177  259  239  264  258  52  67  111  257  94  142  121  191  143  260  265  74  207  101  182  174  124  263  4  172  254  151  246  64  183  76  152  188  261  242  102  34  147  241  134  178  245  176  175  38  162  107  220  65  194  69  247  13  215  78  112  29  139  98  79  158  61  33  106  35  50  16  80  205  161  11  204  37  12  189  9  232  15  228  155  47  171  213  209  218  96  90  212  217  181  163  140  114  115  108  71  141  14  36  48  224  25  105  57  216  138  24  28  211  160  192  43  44  157  154  41  219  8  68  129  75  221  7  10  63  146  45  26  125  193  210  1  60  3  132  81  22  23  208  5  131  135  88  226  6  156  49  91  92  93  201  250  251  72  95  148  39  31  2  30  51  100  255  252  256  109  196  164  170  227  173  83  159  53  137  54  116  118  119  167  168  169  97"
    }
  ]
}