  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "The docking station top cover of the removable patient table may contain a magnetic metal plate.  When docking/undocking the table, the magnetic top cover may become attracted to the magnet.",
      "address_2": "",
      "product_quantity": "8 systems",
      "code_info": "Model # 10432914 and 10432915  Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485",
      "center_classification_date": "20160503",
      "distribution_pattern": "US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.",
      "state": "PA",
      "product_description": "MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.",
      "report_date": "20160511",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1598-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "73947",
      "termination_date": "20170417",
      "more_code_info": "",
      "recall_initiation_date": "20160413",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}