  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98447",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "recall_number": "Z-1590-2026",
      "product_description": "Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP",
      "product_quantity": "125 units",
      "reason_for_recall": "There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.",
      "recall_initiation_date": "20260203",
      "center_classification_date": "20260320",
      "report_date": "20260401",
      "code_info": "HIBC: +B681J02000; E85101LA: IMB-22-013 IMB-24-013 IMB-23-006 IMB-24-001 IMB-24-016 IMB-24-018 IMB-24-007 IMB-21-001 IMB-23-004 IMB-24-015 IMB-22-014 IMB-22-011 IMB-21-008 IMB-22-010 IMB-23-003 IMB-22-002 IMB-16-005 IMB-16-002 IMB-17-007 IMB-21-003 IMB-16-003 IMB-18-006 IMB-16-006 IMB-19-014 IMB-23-007 IMB-19-011 IMB-24-011 IMB-15-001 IMB-24-009 IMB-20-005 IMB-19-009 IMB-15-003 IMB-18-007 IMB-17-011 IMB-18-003 IMB-19-003 IMB-17-006 IMB-20-004 IMB-21-005 IMB-21-006 IMB-20-012 IMB-22-004 IMB-17-008 IMB-20-022 IMB-19-019 IMB-18-004 IMB-17-012 IMB-19-008 IMB-15-004 IMB-16-004 IMB-17-005 IMB-18-002 IMB-18-005 IMB-18-010 IMB-19-005 IMB-15-002 IMB-17-009 IMB-17-010 IMB-24-012 IMB-17-002 IMB-20-016 IMB-22-007 IMB-24-002 IMB-20-010 IMB-22-005 IMB-22-017 IMB-24-021 IMB-19-002 IMB-24-017 IMB-20-021 IMB-19-007 IMB-24-014 IMB-24-004 IMB-24-020 IMB-19-013 IMB-20-009 IMB-24-006 IMB-20-008 IMB-24-023 IMB-24-022 IMB-18-009 IMB-19-001 IMB-19-004 IMB-19-006 IMB-19-010 IMB-19-016 IMB-22-001 IMB-22-012 IMB-22-016 IMB-22-018 IMB-17-003 IMB-24-005 IMB-16-007 IMB-22-015 IMB-24-003 IMB-20-002 IMB-20-003 IMB-20-017 IMB-20-018 IMB-20-019 IMB-21-002 IMB-21-004 IMB-22-003 IMB-22-008 IMB-22-009 IMB-20-011 IMB-20-020; E85101LP:  IMB-25-005 IMB-21-007 IMB-25-007 IMB-25-008 IMB-20-001 IMB-20-013 IMB-25-012 IMB-25-009 IMB-25-006 IMB-21-009 IMB-25-013 IMB-25-010 IMB-19-012 IMB-25-001 IMB-25-004 IMB-24-024 IMB-25-023 IMB-25-003",
      "more_code_info": ""
    }
  ]
}