{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Memphis",
      "address_1": "1800 Pyramid Place",
      "reason_for_recall": "Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.",
      "address_2": "",
      "product_quantity": "22936 units",
      "code_info": "Lots:  M111057AAP, M111064AAD, M111064AAC, M111063AAJ, M111064AAK, M111064AAL, M111064AAV, M111064AAM, M111064AAW, M111064AAG, M111057AAM, M111063AAD, M111064AAQ, M111064AAO, M111064AAP, M111063AAP, M111064AAB, M111063AAC, M111064AAY, M111057AAD, M111057AAE",
      "center_classification_date": "20130625",
      "distribution_pattern": "Worldwide distribution: USA (nationwide including Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, Hong Kong, India, Mexico, Netherlands, , New South Wales, Panama, and Paraguay.",
      "state": "TN",
      "product_description": "INFUSE(R) Bone Graft, SMALL KIT, REF 7510200, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132    orthopaedic",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Sofamor Danek USA Inc",
      "recall_number": "Z-1589-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65297",
      "termination_date": "20140513",
      "more_code_info": "",
      "recall_initiation_date": "20130528",
      "postal_code": "38132",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}