      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse.      Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol",
      "address_2": "",
      "product_quantity": "50 units",
      "code_info": "The following codes are affected by this recall:System Code #728231, Serial #:  9654, 91003, 95130, 95157, 95504    System Code #728326, Serial #:   32000,300003,300004,300005,300010,300011,300012,300013,  300014,300015,300018, 300027,300041,300053,300054, 300059, 300066, 300070,300071, 300073, 300074, 30077, 300078, 300079, 300080,300082,300084,300086,300091,300092, 300095, 300101, 300102,300112, 300119, 300127, 300128, 300129, 300131, 300132, 310064, 320030    System Code #728321, Serial #:  310096, 310070    System Code #728323, Serial #:  320033",
      "center_classification_date": "20130625",
      "distribution_pattern": "Nationwide Distribution-including the states of AR, AZ, DE, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, and WY.",
      "state": "OH",
      "product_description": "Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems.     Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-1588-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65137",
      "termination_date": "20131127",
      "more_code_info": "",
      "recall_initiation_date": "20130321",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}