  "results": [
    {
      "country": "United States",
      "city": "Indianola",
      "address_1": "1 Medrad Dr",
      "reason_for_recall": "The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.",
      "address_2": "",
      "product_quantity": "1,066",
      "code_info": "Catalog numbers 3011992, 3011992W, 3011993, 3011993W, 3011994, 3011994W, 3011996, 3011996W, 3014268 and 3014268W.",
      "center_classification_date": "20130624",
      "distribution_pattern": "Worldwide Distribution -  US Nationwide including the states of  AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and the countries of:  AE, AU, AZ, BN, BR, CA, CH, CL, CO, DE, DK, DO, DZ, EG, ES, FR, GB, HK, HN, HR, HU, IN, IT, JO, JP, KR, LB, MU, MX, MY, NL, PH, PS, SA, SG, TH, TR, TW, UY, VE, and ZA.",
      "state": "PA",
      "product_description": "DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor:     Product Usage:  Monitor, Physiological, Patient",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medrad Inc",
      "recall_number": "Z-1586-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65411",
      "termination_date": "20140403",
      "more_code_info": "",
      "recall_initiation_date": "20130415",
      "postal_code": "15051-9759",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}