  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Richmond",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90336",
      "recalling_firm": "Mckesson Medical-Surgical Inc. Corporate Office",
      "address_1": "9954 Maryland Drive",
      "address_2": "Deep Run Iii Ste. 4000",
      "postal_code": "23233",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-1583-2022",
      "product_description": "a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD.  Part No. 0356-50.      b.  Coagulation Control Kit Coag-Sense PT / INR Test (Prothrombin Time Test) Low Level / High Level 10 Strips Part No. 03P69-10.",
      "product_quantity": "a. 27 boxes  b. 1 box",
      "reason_for_recall": "Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.",
      "recall_initiation_date": "20220525",
      "center_classification_date": "20220825",
      "report_date": "20220831",
      "code_info": "a.  GTIN:  00861900000336; 10861900000395; 00861900000398.  Lot codes distributed between 5/30/2021 - 8/24/2021.    b.  GTIN 00861900000343 Lots Distributed: 8/20/2021 - 8/20/2021",
      "more_code_info": ""
    }
  ]
}