  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Feldkirchen",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70951",
      "recalling_firm": "Brainlab AG",
      "address_1": "Kapellenstrasse 12",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "*** US: Nationwide; *** FOREIGN: Austria, China, Denmark, France, Germany, Hong Kong, Japan, Netherlands, Qatar, Russian Federation, South Korea, Switzerland",
      "recall_number": "Z-1582-2015",
      "product_description": "ExacTrac 6.x. is software used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures.",
      "product_quantity": "25 systems (US only); 53 systems (Foreign)",
      "reason_for_recall": "ExacTrac 6.x Patient Positioning System: Potentially incorrect patient positioning when using the ExacTrac Cone Beam CT (CBCT) with a CBCT acquired at a couch angle other than 0.0 degrees.",
      "recall_initiation_date": "20150306",
      "center_classification_date": "20150506",
      "termination_date": "20170424",
      "report_date": "20150513",
      "code_info": "Software Versions: ExacTrac v. 6.x (v. 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0);     Model/Catalog Numbers: 49971, CBCT IMPORT & ALIGNMENT SOFTWARE; 20836, EXACTRAC 6.0 VARIAN CBCT LICENSE;   48320, ET UPGRADE SOFTWARE 6.0 TO 6.1;   48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA;   49973B, ET DATA PREP / REVIEW SYSTEM;   49973A, ET DATA PREP / REVIEW SYSTEM;   49973, ET DATA PREP / REVIEW SYSTEM;   49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT;   48302, ET DATA PREP / REVIEW SYSTEM"
    }
  ]
}