{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91975",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.",
      "recall_number": "Z-1580-2023",
      "product_description": "DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with breathing circuits",
      "product_quantity": "47007 units",
      "reason_for_recall": "DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.",
      "recall_initiation_date": "20230405",
      "center_classification_date": "20230512",
      "report_date": "20230524",
      "code_info": "Lot Numbers:  3577218, 3738865, 3887812, 4024574, 4151070, 4258733, 3582650, 3752126, 3901443, 4029610, 4153964, 4258734, 3587542, 3761923, 3911867, 4038968, 4164878, 4264897, 3596826, 3768558, 3914813, 4041942, 4173946, 4269655, 3604794, 3768559, 3917564, 4063257, 4176082, 4279829, 3613310, 3772121, 3931273, 4063258, 4182389, 4279830, 3623879, 3779557, 3933681, 4067165, 4187590, 4284606, 3627084, 3784957, 3936715, 4072088, 4192791, 4284607, 3643070, 3788286, 3942733, 4076804, 4201183, 4290269, 3671340, 3798095, 3952706, 4081672, 4204667, 4290270, 3671341, 3802450, 3956054, 4086259, 4215512, 4318561, 3673809, 3805780, 3959458, 4090323, 4219068, 4335149, 3687267, 3819513,  3970976, 4093610, 4229475, 4335150, 3690140, 3828396, 3974773, 4104647, 4242508, 4337436, 3700821, 3836009, 3988375, 4108513, 4242509, 4361916, 3709028, 3849849, 4001786, 4111190, 4247925,  3712265, 3863155, 4005980, 4116119, 4247926, 3723465, 3867378, 4011955, 4132044,  4254339,  3735446, 3880821, 4018725,  4138935, 4258732"
    }
  ]
}